The P940 Pressure
diaphragm is a stainless-steel wetted surface pressure diaphragm
with a proprietary sputtered thin-film full-bridge strain gauge
circuit. The sputtered metal strain gauge circuit offers superior
long term stability and higher temperature capability than other
strain gauge technologies. A unique lead out pad design provides
ease of soldering and rugged construction. The 10KOhm bridge offers
low power consumption for transmitters. Ideal for pressure transducer
OEMs.
* Stainless-Steel Wetted Surface
* Low Power Consumption
* Unsurpassed Stability
* Gold Contact Soldering Pad
* Ultra-High Temperature
Versions Available
* Rugged Construction
* 0.06% FRO Drift/Year
* 10,000 Ohm Full Bridge
* -40°C to 200°C
* 500 Vac Dielectric Available
* 50 Through 6,000 psi Ranges
Click
here to download the specifications
in .pdf format
Specifications
PERFORMANCE
Repeatability
< 0.03 % Span
Over Pressure Limit, Proof
150 % of Range
Over Pressure Limit, Burst
400 % of Range
Long Term Stability
Standard
< 0.8 % Span per year
Optional
< 0.06 % Span per year
Temp. Effect on Zero Balance
Standard
< 0.03 % Span / °C
Optional
< 0.06 % Span / °C
Temp. Effect on Span
Standard
< 0.025 % Reading / °C
Optional
< 0.06 % Reading / °C
Operating Temp. Range
Standard
-40°C (-40°F) to 150°C (302°F)
Optional
-40°C (-40°F) to 200°C
(392°F)
Optional II
Consult factory for higher temps.
Calibrated Temp. Range
-20°C (-4°F) to 125°C (257°F)
Wetted Material
Stainless Steel
ELECTRICAL
Rated Output (R.O.)
2 mV / V nominal
Excitation
5 - 20 V ac or dc
Bridge Configuration
Standard
Full Bridge, 4 wire closed
Optional
Full Bridge, 5 wire open
Bridge Resistance
10,000 Ω ±2 %
Insulation Resistance
Standard
1,000 Meg ohm @ 50 Vdc
Optional
> 500 Vac rms breakdown
Electrical Termination
Gold Solder Pads on Ceramic Substrate
Contact SMD for all of
your
custom application needs
USA:
Strain Measurement Devices
55 Barnes Park North
Wallingford, CT 06492,
USA
P: (203) 294-5800
F: (203) 265-0544 askus@smdsensors.com